CR-150
Description
The CR-150 Stent Crimping Machine is developed for the crimping and transfer into the delivery system of self-expandable stents up to 120 mm length with a maximum diameter of approx. 20 mm.
Specifications
- Automatic stent loading with cradle system
 - Servo controlled stent crimping head with Zero Gap technology
 - Exchangeable pusher rods and delivery system holders for various stent diameters
 - 10” user interface with recipe control
 - The CR-150 is built with a very high-quality benchmark resulting in a stable and robust design with minimal cost of ownership.
 - CR-150 stent crimper can be easily calibrated using the one-point automatic diameter calibration function built into the system. Calibration in seconds.
 - Full process data logging to the internal PC system and transfer possibly to external USB for registration or analysis.
 
Technical Specs
Stent crimping length  | 150 mm head. Maximum 120 mm length stents  | 
Stent type  | Self-expandable stents (also for covered stents)  | 
Maximum diameter auto loading  | 20 mm (optionally larger sizes possible)  | 
Maximum opening crimp head  | 35 mm  | 
Minimum closing diameter  | 0.1 mm (software limit)  | 
Crimp head control  | Servo controlled position / force  | 
Diameter accuracy  | +/- 0.02 mm @ 100 N  | 
Maximum crimp force  | 1000 N radial. Dynamic Diameter Compensation possible  | 
Crimping speeds  | 0.1 mm/s – 10 mm/s  | 
Maximum transfer force  | 200 N with loadcell measurement on the pusher  | 
Stent positioning  | Fully automatic with stent cradle system  | 
GUI  | 10” colour touch panel  | 
Control software  | Zoomable process graphs Recipe based Password levels (operator, engineer, calibrator) Remote support option Data logging with USB transfer possibility  | 
Ethernet  | 2x  | 
USB  | 4x  | 
Required air pressure  | 7 bar minimum  | 
Power supply  | 115 / 240 VAC / 0.5 kW  | 
UPS  | Optional power supply backup to finish current cycle  | 
Safety systems  | Covers with coded magnets and safety PLC system  |